Description
The Honeywell 04230-A-1001 (Type SHC-1) is an intrinsically safe, portable maintenance, configuration, and calibration tool designed for Honeywell Analytics fixed infrared gas detection systems. Running factory-installed 4V0 software, this device serves as the primary field interface for commissioning, interrogating, and diagnostic troubleshooting of Searchpoint Optima, Searchpoint Optima Plus, and Searchline Excel gas detectors. It establishes a digital link with field devices to read operational metrics, execute zero/span calibrations, and download event logs without taking the detectors offline.
Technical Specifications
| Parameter | Value |
| Brand / Manufacturer | Honeywell Analytics (Legacy Zellweger) |
| Part Number / MPN | 04230-A-1001 |
| Model / Type Designation | SHC-1 |
| Software Baseline | Version 4V0 embedded firmware |
| Compatible Detectors | Searchpoint Optima, Searchpoint Optima Plus, Searchline Excel |
| Communication Interface | Digital RS-485 serial link |
| Enclosure Material | Stainless Steel ANSI 316 / Impact-resistant composite |
| Hazardous Area Certification | ATEX Certified (Intrinsically Safe for Zone 1 and Zone 2) |
| Operating Temperature | 0 °C to +40 °C (DMT Evaluated structural envelope: −25 °C to +55 °C) |
| Operating Humidity | 0 to 99% RH (Non-condensing) |
| Net Weight | 0.5 kg (1.1 lbs) |
Supply Status and Procurement Logic
- Condition: New Original / Factory Sealed (New Surplus units occasionally cycled into active rotation)
- Warranty: 12-Month Comprehensive Technical Warranty
- Availability: Active allocation via safety and instrumentation logistics pipelines
- Lead Time: Ships within 1 to 3 business days from order processing
- Pricing: Variable based on international safety component indices. Contact our engineering desk for a firm, contract-specific quote.
Critical Field Integration Rules & Constraints
When using the SHC-1 interrogator on live processing plants, maintenance teams must strictly observe the following constraints to prevent loop failures or code errors:
Certification and Isolation Boundaries
The ATEX intrinsic safety (IS) certification of the 04230-A-1001 interrogator is strictly conditional. It only maintains its safety rating when connected via an approved isolation junction. You must never splice the interrogator lines directly into a bare, non-isolated 24 V DC power loop. Field connection must proceed exclusively via the IS socket on a DVC100 / DX100 termination unit or by deploying the specialized 04230-A-1025 SHC Protection Device module with its factory flying leads.
Regional Certification Variances
The part number 04230-A-1001 is the ATEX-approved model. If your plant operates under alternative regulatory regimes, this exact part number cannot be used as a generic substitute. For projects requiring North American approvals, you must source regional variants: 2104B2351 for UL certification boundaries or 2104B2354 for CSA-regulated jurisdictions.
Firmware-Level Interoperability
The 4V0 software suite embedded within this unit is optimized for the internal register maps of Searchpoint Optima Plus and Searchline Excel cross-duct units. When servicing legacy, non-Plus Searchpoint Optima units, certain advanced extended diagnostic sub-menus may report a null read or communication timeout error. This is an inherent software architecture boundary, not a hardware defect; verify individual detector register maps before executing field-wide parameter alterations.
Frequently Asked Questions (FAQ)
What occurs if the interrogator display screen shows a “No Communication” error upon connection?
Verify that the RS-485 lines are clean and the DVC100/DX100 IS socket pins are free of oxidation. This error usually stems from a baud rate mismatch between the detector’s multi-drop network address configuration and the default interrogation scan routine of the 4V0 software. If the gas detector is actively starting up, wait 2 minutes for its internal diagnostic routine to clear before initiating an interrogation sweep.
Can a zero or span calibration be performed in a classified area using this device?
Yes, because the 04230-A-1001 is certified as intrinsically safe, it allows maintenance technicians to perform full sensor trim, zero calibration, and span alignment directly at the point of detection within a classified hazardous area without requiring a hot work permit.
Is it necessary to recalibrate the interrogator unit itself?
No. The SHC-1 is an electronic interface device that transmits configuration commands and views data registers from the detector; it contains no internal gas-sensing cells. Calibration adjustments are written directly into the non-volatile memory of the target Searchpoint Optima Plus or Searchline Excel device rather than the handheld tool itself. Ensure the unit’s housing seals remain intact to preserve moisture isolation.
| Honeywell 05701-A-0302 Single-Channel Control Card |
| Honeywell 05704-A-0144 System 57 Quad Control Card |
| Honeywell FC-QPP-0002 – Quad Processor Pack |
| Honeywell CC-PWRR01 – Redundant Power Supply Module |